Detailed technical instructions for the BPM (Blood Pressure Monitor) and the Diascan system , which provides real-time Kt/V measurements for dialysis dose monitoring.

| Error Code | Display Message | Likely Cause (per manual) | Corrective Action | |------------|----------------|---------------------------|-------------------| | 088 006 | Flow alarm | Balance chamber membrane rupture | Replace membrane, test chamber integrity | | 088 011 | Conductivity out of range | Concentrate pump calibration drift | Recalibrate N2 and Bicarb pumps | | 088 013 | Temperature alarm | Heater relay stuck or NTC failed | Check heater PCB, replace thermistor | | 088 033 | Blood leak alarm false positive | Air bubbles or dirty sensor | Clean blood leak detector, degas fluid | Gambro Ak 96 Service Manual